Eternity AM834 Anaesthesia Machine

Eternity AM834 Anaesthesia Machine

Product Code:AM834
Availability:In Stock

Specification :

  • Display : 8.4″ TFT color touch screen
  • Monitoring :
    Volumes : MV, VTe, VTI
    Pressures : Ppeak, Pmean, PEEP
    Breath Rate
    RATE spn
    FiO2%
    ETCO2 (Optional)
    P-t, F-t waveforms, P-V and F-V
    Graph display Loops

Pneumatic Specification

  • Gas Supply : O2, N2O, Air and Reserved O2
  • Input Pressure : 0.3MPa~0.6MPa
  • Yoke  :1 x O2, 1 x N2O (Optional)
  • Flowmeter : O2 :0~1L/min, 1~10L/min | N2O :0~1L/min, 1~10L/min | AIR :0~1L/min, 1~10L/min
  • Pressure gauge : 3for pipeline gas, 2 for cylinder gas (Optional)
  • O2 flush : 25~75L/min
  • Operational Modes : Closed and semi-closed circuit system, semi open
  • Driven mode : Pneumatically driven and electronically controlled
  • AGSS (Passive) : Optional
  • Gas Module (Optional) : CO2, AG

Breathing System

  • APL Valve : 0.19~6KPa
  • CO2 canister : 1800ml (Single Canister)
  • Compensation for circuit leakage and compliance
  • Bypass Function :  (Optional)

Physical Specifications

  • Dimension : L880 * W800 * H1580 mm
  • Weight : 120kg

Electronic Specification

  • Power Supply : AC100-240V±10%, 50/60Hz±1%
  • Battery Backup : Minimum 120min
  • Power Consumption :  ≤70VA
  • Top LED light
  • RJ45 (optional)

Alarm

  • High airway pressure / Low airway pressure
  • High tidal volume / Low tidal volume
  • High frequency / Low frequency
  • High minute volume / Low minute volume
  • High FiO2 / Low FiO2 (should option FiO2 function)
  • Apnea
  • Power Failure
  • Low Battery
  • Low Gas Supply
  • Alarm Silence ≤120 seconds

Vaporizer

  • Two vaporizer position (Selecta Tec with interlock)
  • Optional two vaporizer positions
  • Agent type : Sevoflurane, Halothane, Isoflurane and Enflurane
  • Filling type : Pour-fill, Easy-fill, Quick-fill or Key-fill
  • Compensation : Flow, pressure and temperature compensation

Ventilator Specification

  • Patient Type : Pediatric and adult
  • Ventilation Mode : VCV, PCV, SIMV, STANDBY, Manual
  • Tidal Volume : 15-1500ml
  • Frequency : 0-100bpm, SIMV frequency : 4-20bpm
  • I:E ratio : 10:1 – 1:10
  • Inspiratory Pause : 0-60% of inspiratory time
  • Inspiratory Time : 0.5 – 5S
  • Pmax : 5-70cmH2O
  • PEEP : OFF, 4-30cmH2O
  • Pressure Trigger : 0-20cmH2O
  • Flow Trigger : 0-20L/min